<?xml version="1.0" encoding="UTF-8"?>
<!-- edited with XMLSpy v2008 rel. 2 (http://www.altova.com) by Paul Schluter (GE Healthcare) -->
<partition xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns="http://www.ieee.org/11073/nomenclature" xsi:schemaLocation="http://www.ieee.org/11073/nomenclature terms.8f.xsd">
	<!-- Edited and verified against ecg_nom5c.doc PSS 2008-05-05 -->
	<!-- Final code assignments for 10 new terms starting at 2::32768, using 16*256 block allocated by Jan.  PSS 2008-08-31 -->
	<!-- Should consider reserving 32*256 rather than 16*256 codes if possible. -->
	<!-- perLead offset of 0 is "global"? -->
	<!-- 2009-04-30 Edited/transformed to conform to terms.8f.xsd (harmonized with IDCO and RTM) -->
	<partitionDescription>
		<title>ECG Measurements, per Lead</title>
		<date>2012-09-26</date>
		<version>0.8.7</version>	<!-- "0.8.7" corresponds to "8g", updated on 2009-09-26  -->
		<discriminatorRef idref="MDC_ECG_discrim_LEADS"/>
		<!-- basePart is not used since each <term> conveys the partition -->
		<baseCode>0</baseCode>
	</partitionDescription>
	<!-- -->
	<Terms>
		<term>
			<SysName>Duration | ECG {lead}, P | Heart | CVS</SysName>
			<DisplayName>P duration</DisplayName>
			<Description>Duration of the P wave of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_P</REFID>
			<PART>2</PART>
			<CODE10>6656</CODE10>
		</term>
		<term>
			<SysName>Timepoint | ECG {lead}, P, End | Heart | CVS</SysName>
			<Mnemonic>Poff</Mnemonic>
			<Description>Time point of end of P wave in a specified {lead}</Description>
			<REFID>MDC_ECG_TIME_END_P</REFID>
			<PART>2</PART>
			<CODE10>5888</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, P, FirstExtremum | Heart | CVS</SysName>
			<Description>Duration of the interval between P onset and the first extremum of the P wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_P1</REFID>
			<PART>2</PART>
			<CODE10>4608</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, P, SecondExtremum | Heart | CVS</SysName>
			<Description>Duration of the interval between P onset and the second extremum of the P wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_P2 </REFID>
			<PART>2</PART>
			<CODE10>4864</CODE10>
		</term>
		<term>
			<SysName>Timepoint | ECG {lead}, P, Start | Heart | CVS</SysName>
			<Mnemonic>Pon</Mnemonic>
			<Description>Time point of start of P wave in a specified {lead}</Description>
			<REFID>MDC_ECG_TIME_START_P</REFID>
			<PART>2</PART>
			<CODE10>9472</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, P, ThirdExtremum | Heart | CVS</SysName>
			<Description>Duration of the interval between P onset and the third extremum of the P wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_P3 </REFID>
			<PART>2</PART>
			<CODE10>5120</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, PR | Heart | CVS</SysName>
			<DisplayName>P-R interval</DisplayName>
			<Mnemonic>PR</Mnemonic>
			<Description>Duration of the interval between P offset and QRS onset of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_PR</REFID>
			<PART>2</PART>
			<CODE10>7168</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, Q | Heart | CVS</SysName>
			<DisplayName>Q wave duration</DisplayName>
			<Description>Duration of the Q wave of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_Q</REFID>
			<PART>2</PART>
			<CODE10>7680</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, QRS | Heart | CVS</SysName>
			<DisplayName>QRS duration</DisplayName>
			<Description>Duration of the QRS complex of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_QRS</REFID>
			<PART>2</PART>
			<CODE10>7936</CODE10>
		</term>
		<term>
			<SysName>Timepoint | ECG {lead}, QRS, End | Heart | CVS</SysName>
			<Mnemonic>QRSoff</Mnemonic>
			<Description>Time point of end of QRS complex in a specified {lead}</Description>
			<REFID>MDC_ECG_TIME_END_QRS</REFID>
			<PART>2</PART>
			<CODE10>6144</CODE10>
		</term>
		<term>
			<SysName>Timepoint | ECG {lead}, QRS, Start | Heart | CVS</SysName>
			<Mnemonic>QRSon</Mnemonic>
			<Description>Time point of start of QRS complex in a specified {lead}</Description>
			<REFID>MDC_ECG_TIME_START_QRS</REFID>
			<PART>2</PART>
			<CODE10>9728</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, QT | Heart | CVS</SysName>
			<DisplayName>Q-T interval</DisplayName>
			<Mnemonic>QT</Mnemonic>
			<Description>Duration of the interval between the QRS onset and T wave offset of ECG in {lead} (used for QT dispersion)</Description>
			<REFID>MDC_ECG_TIME_PD_QT</REFID>
			<PART>2</PART>
			<CODE10>8192</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, QTc | Heart | CVS</SysName>
			<DisplayName>Q-T c</DisplayName>
			<Mnemonic>QTc</Mnemonic>
			<Description>Duration of the interval between the QRS onset and T wave offset, related to heart rate 60 beats per minute of ECG in {lead}, Bazett formula </Description>
			<REFID>MDC_ECG_TIME_PD_QT_CORR</REFID>
			<PART>2</PART>
			<CODE10>8448</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, R1 | Heart | CVS</SysName>
			<DisplayName>R1 wave duration</DisplayName>
			<Mnemonic>R1</Mnemonic>
			<Description>Duration of the R1 wave of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_R_1</REFID>
			<PART>2</PART>
			<CODE10>11264</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, R2 | Heart | CVS</SysName>
			<DisplayName>R2 wave duration</DisplayName>
			<Mnemonic>R2</Mnemonic>
			<Description>Duration of the R2 wave of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_R_2 </REFID>
			<PART>2</PART>
			<CODE10>11520</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, R3 | Heart | CVS</SysName>
			<DisplayName>R3 wave duration</DisplayName>
			<Mnemonic>R3</Mnemonic>
			<Description>Duration of the R3 wave of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_R_3</REFID>
			<PART>2</PART>
			<CODE10>11776</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, S1 | Heart | CVS</SysName>
			<DisplayName>S1 wave duration</DisplayName>
			<Mnemonic>S1</Mnemonic>
			<Description>Duration of the S1 wave of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_S_1 </REFID>
			<PART>2</PART>
			<CODE10>12032</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, S2 | Heart | CVS</SysName>
			<DisplayName>S2 wave duration</DisplayName>
			<Mnemonic>S2</Mnemonic>
			<Description>Duration of the S2 wave of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_S_2</REFID>
			<PART>2</PART>
			<CODE10>12288</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, S3 | Heart | CVS</SysName>
			<DisplayName>S3 wave duration</DisplayName>
			<Mnemonic>S3</Mnemonic>
			<Description>Duration of the S3 wave of ECG in {lead}</Description>
			<REFID>MDC_ECG_TIME_PD_S_3</REFID>
			<PART>2</PART>
			<CODE10>12544</CODE10>
		</term>
		<term>
			<SysName>Timepoint | ECG {lead}, T, End | Heart | CVS</SysName>
			<Mnemonic>Toff</Mnemonic>
			<Description>Time point of end of T wave in a specified {lead}</Description>
			<REFID>MDC_ECG_TIME_END_QRS</REFID>
			<PART>2</PART>
			<CODE10>6400</CODE10>
		</term>
		<term>
			<SysName>Timepoint | ECG {lead}, T, Start | Heart | CVS</SysName>
			<Mnemonic>Ton</Mnemonic>
			<Description>Time point of start of T wave in a specified {lead}</Description>
			<REFID>MDC_ECG_TIME_START_T</REFID>
			<PART>2</PART>
			<CODE10>9984</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, VentricularActivation | Heart | CVS</SysName>
			<DisplayName>Ventricular Activation Time</DisplayName>
			<Mnemonic>VAT</Mnemonic>
			<Description>Ventricular activation time</Description>
			<REFID>MDC_ECG_TIME_PD_VENT_ACTIV</REFID>
			<PART>2</PART>
			<CODE10>11008</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead} | Heart | CVS</SysName>
			<DisplayName>ECG {lead code}</DisplayName>
			<Mnemonic>ECG-{lead}</Mnemonic>
			<Description>ECG as recorded according to {lead} in specified position (time series)</Description>
			<REFID>MDC_ECG_ELEC_POTL</REFID>
			<PART>2</PART>
			<CODE10>256</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, J | Heart | CVS</SysName>
			<Mnemonic>ST-J</Mnemonic>
			<Description>Amplitude at the end of QRS complex (junctional point or J point) of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_AMPL_J</REFID>
			<PART>2</PART>
			<CODE10>1024</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, J20 | Heart | CVS</SysName>
			<Mnemonic>ST-J20</Mnemonic>
			<Description>Amplitude at 20 ms after the end of QRS complex of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_ST_20</REFID>
			<PART>2</PART>
			<CODE10>14848</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, J40 | Heart | CVS</SysName>
			<Mnemonic>ST-J40</Mnemonic>
			<Description>Amplitude at 40 ms after the end of QRS complex of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_ST_40</REFID>
			<PART>2</PART>
			<CODE10>15104</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, J60 | Heart | CVS</SysName>
			<Mnemonic>ST-J60</Mnemonic>
			<Description>Amplitude at 60 ms after the end of QRS complex of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_ST_60</REFID>
			<PART>2</PART>
			<CODE10>14336</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, J80 | Heart | CVS</SysName>
			<Mnemonic>ST-J80</Mnemonic>
			<Description>Amplitude at 80 ms after the end of QRS complex of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_ST_80</REFID>
			<PART>2</PART>
			<CODE10>14592</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, P, FirstExtremum | Heart | CVS</SysName>
			<DisplayName>Maximum P wave amplitude</DisplayName>
			<Mnemonic>Pmax</Mnemonic>
			<Description>Amplitude level of first extremum of the P wave of ECG in specified {lead} (mostly maximum of P, depends on morphology)</Description>
			<REFID>MDC_ECG_AMPL_P_MAX</REFID>
			<PART>2</PART>
			<CODE10>1280</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, P, SecondExtremum | Heart | CVS</SysName>
			<DisplayName>Minimum P wave amplitude</DisplayName>
			<Mnemonic>Pmin</Mnemonic>
			<Description>Amplitude level of the second extremum of the P wave of ECG in specified {lead} (often minimum of P, depends on morphology)</Description>
			<REFID>MDC_ECG_AMPL_P_MIN</REFID>
			<PART>2</PART>
			<CODE10>1536</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, P, ThirdExtremum | Heart | CVS</SysName>
			<Mnemonic>P3</Mnemonic>
			<Description>Amplitude level of the third extremum of the P wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_AMPL_P3</REFID>
			<PART>2</PART>
			<CODE10>3072</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, Q | Heart | CVS</SysName>
			<DisplayName>Q wave amplitude</DisplayName>
			<Description>Amplitude of the Q wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_AMPL_Q </REFID>
			<PART>2</PART>
			<CODE10>1792</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, R, Maximum | Heart | CVS</SysName>
			<Mnemonic>Rmax</Mnemonic>
			<Description>Maximum amplitude of R wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_AMPL_R</REFID>
			<PART>2</PART>
			<CODE10>2048</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, R1 | Heart | CVS </SysName>
			<DisplayName>R1 wave amplitude</DisplayName>
			<Description>Amplitude of the R1 wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_R_1</REFID>
			<PART>2</PART>
			<CODE10>12800</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, R2 | Heart | CVS </SysName>
			<DisplayName>R2 wave amplitude</DisplayName>
			<Description>Amplitude of the R2 wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_R_2</REFID>
			<PART>2</PART>
			<CODE10>13056</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, R3 | Heart | CVS </SysName>
			<DisplayName>R3 wave amplitude</DisplayName>
			<Description>Amplitude of the R3 wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_R_3</REFID>
			<PART>2</PART>
			<CODE10>13312</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, S, Maximum | Heart | CVS</SysName>
			<Mnemonic>Smax</Mnemonic>
			<Description>Maximum amplitude of S wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_AMPL_S</REFID>
			<PART>2</PART>
			<CODE10>2304</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, S1 | Heart | CVS</SysName>
			<DisplayName>S1 wave amplitude</DisplayName>
			<Description>Amplitude of the S1 wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_S_1</REFID>
			<PART>2</PART>
			<CODE10>13568</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, S2 | Heart | CVS</SysName>
			<DisplayName>S2 wave amplitude</DisplayName>
			<Description>Amplitude of the S2 wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_S_2 </REFID>
			<PART>2</PART>
			<CODE10>13824</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, S3 | Heart | CVS</SysName>
			<DisplayName>S3 wave amplitude</DisplayName>
			<Description>Amplitude of the S3 wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_ELEC_POTL_S_3</REFID>
			<PART>2</PART>
			<CODE10>14080</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, T, Maximum | Heart | CVS</SysName>
			<DisplayName>Positive T wave amplitude</DisplayName>
			<Mnemonic>Tmax</Mnemonic>
			<Description>Amplitude of the T positive wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_AMPL_T_MAX</REFID>
			<PART>2</PART>
			<CODE10>2560</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG {lead}, T, Minimum | Heart | CVS</SysName>
			<DisplayName>Negative T wave amplitude</DisplayName>
			<Mnemonic>Tmin</Mnemonic>
			<Description>Amplitude of the T negative wave of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_AMPL_T_MIN</REFID>
			<PART>2</PART>
			<CODE10>2816</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG, lead set | Heart | CVS</SysName>
			<DisplayName>ECG, Lead set</DisplayName>
			<Mnemonic>ECG</Mnemonic>
			<Description>ECG as recorded from a set of leads (time sample array is unspecified lead as an composite element of specified or unspecified leads)</Description>
			<REFID>MDC_ECG_ELEC_POTL</REFID>
			<PART>2</PART>
			<CODE10>256</CODE10>
		</term>
		<term>
			<SysName>ElectricalPotential | ECG{lead}, ST | Heart | CVS</SysName>
			<DisplayName>STTxx Amplitude</DisplayName>
			<Mnemonic>ST-Jxx</Mnemonic>
			<Description>Amplitude of the ST segment at xx ms after the end of QRS complex of ECG in specified {lead}. The time point xx is defined globally by item with code XXXXXX</Description>
			<REFID>MDC_ECG_AMPL_ST</REFID>
			<PART>2</PART>
			<CODE10>768</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, P | Heart | CVS</SysName>
			<DisplayName>P wave integral</DisplayName>
			<Mnemonic>Pintegral</Mnemonic>
			<Description>Integral of the P wave of ECG in specified {lead} (mVolt x milisecond)</Description>
			<REFID>MDC_ECG_INTEGRAL_P</REFID>
			<PART>2</PART>
			<CODE10>6912</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, P, Area | Heart | CVS</SysName>
			<DisplayName>P wave area</DisplayName>
			<Mnemonic>Parea</Mnemonic>
			<Description>Area of the P wave of ECG in specified {lead} (mVolt x millisecond) by integrating absolute values.</Description>
			<REFID>MDC_ECG_AREA_P</REFID>
			<PART>2</PART>
			<CODE10>3840</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, Q | Heart | CVS</SysName>
			<DisplayName>Q wave integral</DisplayName>
			<Mnemonic>Qintegral</Mnemonic>
			<Description>Integral of the Q wave of ECG in specified {lead} (mVolt x millisecond)</Description>
			<REFID>MDC_ECG_INTEGRAL_Q </REFID>
			<PART>2</PART>
			<CODE10>7424</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, Q, Area | Heart | CVS</SysName>
			<DisplayName>Q wave area</DisplayName>
			<Mnemonic>Qarea</Mnemonic>
			<Description>Area of the Q wave of ECG in specified {lead} (mVolt x millisecond) by integrating absolute values.</Description>
			<REFID>MDC_ECG_AREA_Q</REFID>
			<PART>2</PART>
			<CODE10>3328</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, QRS | Heart | CVS</SysName>
			<DisplayName>QRS integral</DisplayName>
			<Mnemonic>QRSintegral</Mnemonic>
			<Description>Integral of the QRS complex of ECG in specified {lead} (mVolt x millisecond)</Description>
			<REFID>MDC_ECG_INTEGRAL_QRS</REFID>
			<PART>2</PART>
			<CODE10>8704</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, QRS, Area | Heart | CVS</SysName>
			<DisplayName>QRS area</DisplayName>
			<Mnemonic>QRSarea</Mnemonic>
			<Description>Area of the QRS complex of ECG in specified {lead} (mVolt x millisecond) by integrating absolute values.</Description>
			<REFID>MDC_ECG_AREA_QRS</REFID>
			<PART>2</PART>
			<CODE10>4096</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, ST | Heart | CVS</SysName>
			<DisplayName>ST-T integral</DisplayName>
			<Mnemonic>ST-Tintegral</Mnemonic>
			<Description>Integral of the ST-T segment computed between J point and the beginning of the T wave of ECG in specified {lead} (mVolt x millisecond)</Description>
			<REFID>MDC_ECG_INTEGRAL_ST</REFID>
			<PART>2</PART>
			<CODE10>9216</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, ST, Area | Heart | CVS</SysName>
			<DisplayName>ST-T area</DisplayName>
			<Mnemonic>ST-Tarea</Mnemonic>
			<Description>Area of the ST-T segment computed between J point and the beginning of the T wave of ECG in specified {lead} (mVolt x millisecond) by integrating absolute values.</Description>
			<REFID>MDC_ECG_AREA_ST</REFID>
			<PART>2</PART>
			<CODE10>4352</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, T | Heart | CVS</SysName>
			<DisplayName>T wave integral</DisplayName>
			<Mnemonic>Tintegral</Mnemonic>
			<Description>Integral of the T wave of ECG in specified {lead} (mVolt x millisecond)</Description>
			<REFID>MDC_ECG_INTEGRAL_T</REFID>
			<PART>2</PART>
			<CODE10>8960</CODE10>
		</term>
		<term>
			<SysName>Integral | ECG {lead}, T, Area | Heart | CVS</SysName>
			<DisplayName>T wave area</DisplayName>
			<Mnemonic>Tarea</Mnemonic>
			<Description>Area of the T wave of ECG in specified {lead} (mVolt x millisecond) by integrating absolute values.</Description>
			<REFID>MDC_ECG_AREA_T</REFID>
			<PART>2</PART>
			<CODE10>3584</CODE10>
		</term>
		<term>
			<SysName>Slope | ECG {lead}, QRS_offset_+_20ms_to_QRS_offset_+_60ms | Heart | CVS</SysName>
			<Description>Slope of ST-segment between QRS offset plus 20 ms and QRS offset plus 60 ms of ECG in specified {lead}</Description>
			<REFID>MDC_ECG_SLOPE_ST </REFID>
			<PART>2</PART>
			<CODE10>5376</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, PP | Heart | CVS</SysName>
			<DisplayName>P-P interval</DisplayName>
			<Mnemonic>PP</Mnemonic>
			<Description>Time period, PP</Description>
			<REFID>MDC_ECG_TIME_PD_PP</REFID>
			<PART>2</PART>
			<CODE10>32768</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, RR | Heart | CVS</SysName>
			<DisplayName>R-R interval</DisplayName>
			<Mnemonic>RR</Mnemonic>
			<Description>Time period, RR</Description>
			<REFID>MDC_ECG_TIME_PD_RR</REFID>
			<PART>2</PART>
			<CODE10>33024</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, PQ | Heart | CVS</SysName>
			<DisplayName>P-Q interval</DisplayName>
			<Mnemonic>PQ</Mnemonic>
			<Description>Time period, PQ</Description>
			<REFID>MDC_ECG_TIME_PD_PQ</REFID>
			<PART>2</PART>
			<CODE10>33280</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, PQ segment | Heart | CVS</SysName>
			<DisplayName>P-Q segment</DisplayName>
			<Mnemonic>Pqseg</Mnemonic>
			<Description>Time period, PQ segment (?)</Description>
			<REFID>MDC_ECG_TIME_PD_PQ_SEG</REFID>
			<PART>2</PART>
			<CODE10>33536</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, QTc | Heart | CVS</SysName>
			<DisplayName>Q-Tc</DisplayName>
			<Mnemonic>QTc</Mnemonic>
			<Description>Time period, QT (correction specified as a separate attribute or unknown)</Description>
			<REFID>MDC_ECG_TIME_PD_QTc</REFID>
			<PART>2</PART>
			<CODE10>33792</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, QTcB | Heart | CVS</SysName>
			<DisplayName>Q-TcB</DisplayName>
			<Mnemonic>QTcB</Mnemonic>
			<Description>Time period, QT (corrected) using Bazett’s correction formula</Description>
			<REFID>MDC_ECG_TIME_PD_QTcB</REFID>
			<PART>2</PART>
			<CODE10>34048</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, QTcF | Heart | CVS</SysName>
			<DisplayName>Q-TcF</DisplayName>
			<Mnemonic>QTcF</Mnemonic>
			<Description>Time period, QT (corrected) using Fredericia's correction formula</Description>
			<REFID>MDC_ECG_TIME_PD_QTcF</REFID>
			<PART>2</PART>
			<CODE10>34304</CODE10>
		</term>
		<term>
			<SysName>Duration | ECG {lead}, QTU | Heart | CVS</SysName>
			<DisplayName>Q-TU</DisplayName>
			<Mnemonic>QTU</Mnemonic>
			<Description>Time period, QTU</Description>
			<REFID>MDC_ECG_TIME_PD_QTU</REFID>
			<PART>2</PART>
			<CODE10>34560</CODE10>
		</term>
		<term>
			<SysName>Shape | ECG {lead}, ST segment after J-point | Heart | CVS</SysName>
			<Description>ST segment shape immediately after J-point<footnote>SHAPE_ST and SHAPE_ST_T can have the following enumerated values:  “concave_up”, “linear”, “concave_down” and “unknown”.</footnote>
				<footnote>At the present time, there is no standard way of expressing the curvature of the ST segment as a numeric parameter.  Mathematically, “curvature” can be defined as the radius of a best-fit circle that is tangent to the area of interest on the ST or ST-T segment on its concave side, using the convention of +|1/R| for “concave_up”, 0 for “linear” and -|1/R| for “concave_down”.  Although there is no experimental or theoretical evidence to support the use of this or any other measure of curvature for the ST and ST-T segment, it is visually and mathematically intuitive.  A standard “visual” unit of measure such as “1mm” (40msec by 100μV) could be assumed.  Other measures are possible (e.g. best-fit polynomial coefficients) and should be considered as the basis for defining two new variables, CURVATURE_ST and CURVATURE_ST_T.  A key benefit of defining a numeric parameter is that new criteria can be retrospectively applied.</footnote>
			</Description>
			<REFID>MDC_ECG_SHAPE_ST</REFID>
			<PART>2</PART>
			<CODE10>34816</CODE10>
		</term>
		<term>
			<SysName>Shape | ECG {lead}, ST segment after ST-T inflection | Heart | CVS</SysName>
			<Description>Second ST-T segment shape (after ST-T inflection point)</Description>
			<REFID>MDC_ECG_SHAPE_ST_T</REFID>
			<PART>2</PART>
			<CODE10>35072</CODE10>
		</term>
	</Terms>
</partition>